CTC PAL Control for Waters Emp3 CDS Installation Manual 3. SUPPORTED CONFIGURATIONS The software is designed for use with Empower 3 Build 3471 Feature Release 2 and higher (for more details see 4.2 Tested Configurations). Contacting Waters Contact Waters ® with enhancement requests or technical quest ions regarding the use, transportation, removal, or disposal of any Waters product. You can reach us via the Internet, telephone, or conventional mail. Empower 3 software Intended use Use the Waters Empower 3 software for acquiring, processing, reporting,.


Tip #156: Creating Simple Summary Reports in Empower Software

Waters Empower 3 Manual

Welcome back to Get Empowered! In the last Empower tip-of-the-week post for Empower Chromatography Data Software, we learned about different types of summary reports (Tip #155).

In this week’s tip, we will learn how to create a simple summary report summarizing replicate injections of a standard. Many SOPs require a %RSD for peak area on replicate injections of standards or controls and a report with an overlay of the chromatograms. There are several preconfigured report groups that make creating the report very straight forward.

Let’s get started.

Step 1

Highlight ‘Results’, right click and select ‘Preview/Publisher’ (figure 1).

Step 2

Select ‘Use the currently open Report Method named Untitled’ from the Open Report Method dialogue and click ‘OK’. Right click on the Main Page view, select ‘Method Properties’ and select ‘Summary By All’ and click ‘OK’ (figure 2).

Figure 2

Step 3

Expand ‘Chromatograms’ in left sidebar and select one of the overlay chromatogram groups (figure 3).

Step 4

The chromatogram properties can be modified depending on preference. From the ‘Plot’ tab, overly the chromatograms in a single plot or ‘Make Stack Plot’. (Note: I recommend annotating ‘All chromatograms in Overlay’). Modify the legend on the ‘Legend’ tab (figure 4).

Figure 4

Step 5

Expand ‘Tables’ in left sidebar and select one of the summary tables. In this example, I selected ‘Area Component Summary’ which will provide statistical calculations for the replicate injections (figure 5).

Step 6

In the properties of the table, go to the ‘Component Summary’ tab. This table is setup to provide mean, standard deviation, and percent relative deviation for area. The displayed precision can be modified as well (figure 6).

Figure 6

Step 7

Center the groups by highlighting both. In the ‘Layout’ menu, select ‘Align Controls’ > ‘Center’ (figure 7).

Step 8

Manual

Be sure to add a header and footer to the report (Refer to Tip #154), save all changes and click ‘Print Preview’ (figure 8).

Figure 8

It’s that easy!

Final Note: This procedure can be followed using the QuickStart or Pro interface.

Please rate this Empower Tip of the Week

Next week in Empower Tip #157 – Next week we will continue our series on Report Methods and learn how to create a slightly more complex summary report.

More Resources:

  • Get the latest Empower projects, options, and more. Check out the New Waters Marketplace.
  • Want to learn more? Check out our Empower training courses and eLearning.
  • Minimize your lab downtime and protect your business against potentially expensive disruptions with the Empower Enterprise Software Extended Critical Support Plan with 24 x 7 x 365 coverage – Learn more
  • Build your technical skills and professional network. Share, ask, and interact with peers from around the globe in dynamic discussions by joining our Informatics Community Forum.
  • Empower is the first cloud-deployable, compliance-ready, enterprise chromatography data software – Learn more about Empower Cloud
Waters Empower 3 Manual

Do you want Empower Tips sent to your inbox every week?

Questions? Tips of your own? Let us know!

Tips on Implementing, Maintaining and Validating Empower 3 Software Custom Fields to Maximize Data Integrity

Waters Empower 3 software is a chromatography data system (CDS) that links to chromatographic instruments to help facilitate management of chromatography test results through data acquisition, processing, reporting and distribution. Waters Empower 3 generates lots of data with complex calculations and detailed reports and has several tools that address compliance and data integrity concerns such as electronic signatures and audit trails.

Given that Waters Empower 3 typically generates large amounts of data, transcribing the data from Empower 3 to an external system can present huge risks from a data integrity perspective. As a result, scientific laboratories are looking for ways to maximize their results output from Empower and ideally utilize this system as their only solution for chromatographic results.

Empower is one of the most versatile and complete Chromatographic Data Systems (CDS) on the market and is able to handle multiple calculation requirements with extensive default functionalities. Often, however, additional customized calculations using “custom fields” are required to meet all the results requirements of modern scientific laboratories. In this blog, we will discuss best practices for implementing and validating Empower custom fields in order to minimize data integrity risk in your laboratory.

Waters empower 3 manual pdf download

Implementing Custom Fields for Waters Empower 3

Custom Fields are a combination of input fields and calculations that allow the user to input minimal data and receive an output of a calculated value utilizing Empower database field data. Before utilizing custom fields to generate your required results, however, first check whether the calculation requirements specified in a given test method or procedure could be accomplished by using default functionalities of the system (e.g., Amount, %Area, RT Ratio, etc…). If the default functionalities are not sufficient, then creating custom fields will be necessary.

The most common types of Custom Fields are “Text” or “Calculated”.

  • Text Custom Fields are usually used to input values that Empower does not have the ability to natively receive (e.g., number of tablets, composite weight, label claim, recovery factors, etc…). It is important to create these fields together with a procedure/work instruction of how to use them and what to enter in each one of them so as to limit the user’s creativity and bring consistency to how the results are being generated between all the users in the Laboratory.
  • Calculated Custom Fields are used to perform calculations when Empower default functionalities are not able to meet the test method/procedure requirements. These Custom Fields can automate the calculation process, remove the necessity of data transcription to an external system (such as an Excel spreadsheet), and reduce the review time required for a result release.

A few key best practices when working with Custom Fields follow:

Use Generic Names. When creating new custom fields, it is advisable to keep their names as generic as possible in order to allow them to be leveraged and reused in other applicable calculations. For example, naming a custom field “Percent_Content” allows you to reuse the field much easier and more often than if it was named “Percent_Ethanol”.

Set Up Appropriate User Privileges. To be able to create Custom Fields, you need to have the correct privileges in the System. Empower has 5 privileges specific for Custom Fields: Create, Alter, Delete, Lock and Unlock. You must also have the “Alter Any Project” privilege enabled in the user type to create a custom field.

Utilize All-In-One Reports. Custom fields also give you the ability to build All-In-One Reports that have all the information needed to properly assess a chromatographic analysis, while also being flexible enough to be used on runs from one to several samples. All-in-One Reports are easier to deal with, as multiple reports create higher volume of documents to review and increase the likelihood of reviewers missing something. In addition, having a locked All-in-One Report restricts user from modifying column headers, assures that the report is built for intended use, and reduces review cycle due to its standardized template.

Create a User Guide Document. When implementing Custom Fields, a user guide document should be created that contains all the information necessary for the Custom Fields to function as intended – how to set up sample sets, processing methods, etc. A User Guide is a necessity for the general user, as Custom Fields require standardized inputs in order to function properly.

Benefits of Implementing Custom Fields and All-in-One Reports

Implementing Custom Fields and All-In-One Reports allows more data to be treated and reviewed electronically, which will provide a number of important benefits for your organization:

  • Quality Control– significant reduction in transcription errors, investigations and data integrity observations related to Empower.
  • Saved Time – reduction in time spent on review cycle, report creation, user training, test method orientation, etc.
  • Operational Efficiency – increases in the number of products tested and reported per analyst in Empower, along with number of tested products reviewed by reviewer in Empower.

Validating Custom Fields for Waters Empower 3

Although Custom Fields functionality could have been tested during the Empower Operational Qualification, there is still a need to validate each Custom Field to provide evidence that the custom field being used is the correct one as per test method requirements. Some best practice recommendations follow:

Create an SOP. As part of the validation strategy, a SOP (Standard Operational Procedure) that describes the procedures to follow when developing, qualifying, implementing and maintaining Custom Fields within Empower should be created.

Create Good Documentation. As part of the validation, a protocol that references the User Guide should be created for the developed Custom Fields. In addition, manual calculation comparison documentation should be created to document the entire cycle of test method-based Custom Fields and report development, review and validation.

Leverage Previously Validated Fields. Validation of Custom Fields can leverage previously validated native fields and previously validated Custom Fields, allowing the implementation of those Custom Fields with minimum testing.

Waters Empower 3 Software Manual

Follow Change Control Processes. When a Test Method that has associated Custom Fields is changed after validation, an impact assessment must be performed within a change control. If Custom Fields are impacted by a Test Method change, the Custom Fields need to be adjusted accordingly, with all adjustments following change control documentation and validation that is commensurate with the change.

Waters Empower 3 Manual Pdf

Conclusion

Custom fields are a great tool to minimize any Data Integrity concerns that transcribing data into external systems poses, while at the same time providing huge gains in operational efficiency. By following the best practice recommendations described in this blog, the development and validation of Custom Fields can be done quickly, resulting in a process that is much less time-consuming, more accurate, less prone to errors, easily traceable and more user-friendly.

Waters Empower 3 User Manual Pdf

Astrix Technology Group has been helping scientific organizations implement and integrate laboratory informatics systems since 1995. The Astrix Team contains highly experienced and knowledgeable Waters Empower 3 experts that can help your organization implement innovative solutions that turn data into knowledge, increase organizational efficiency, improve quality and facilitate regulatory compliance.

The Astrix Difference:

Waters Empower 3 Software User Guide

  • Over 3,500 implemented and tested Custom Fields
  • We understand how to implement Custom Fields based on user needs (e.g., Result set only, Result set first, All or Nothing, Missing Peaks, etc.)
  • We help our clients create a template document for standardizing user input on Sample Set, Processing Method, and Amount Tab
  • We have implemented All-In-One Reports for many clients
  • We can provide a complete validation package that has been utilized at many client implementations (Includes but not limited to: SOP, Protocol Template, Manual Calculation Documentation Template, and documentation to implement the User Input Standardization Document)